Scope and object
This clause of Part 1 is applicable except as follows:
Equipment included in scope Replacement:
This part of IEC 61010 specifies safety requirements for
electrically powered laboratory equipment for the heating of
materials, where the heating of materials is one of the functions
of the equipment.
NOTE If all or part of the equipment falls within the scope of
one or more other part 2 standards of IEC 61010 as well as within
the scope of this standard, it will also need to meet the
requirements of those other part 2 standards. In particular, if
equipmen is intended to be used for IVD purposes, it will need to
meet the requirements of IEC 61010-2-101.
Equipment excluded from scope Addition after item j):
aa) equipment for the heating and ventilation of
laboratories;
bb) sterilizing equipment;
cc) heating and/or cooling equipment which the OPERATOR is
intended to enter, and which is large enough for the OPERATOR to
remain inside with the door or doors closed.